Frequently Asked Questions

Q-SUN Xenon Arc Test Chambers

Q: Can the Q-SUN Run “Solar Radiation” Tests Found in Environmental Testing Standards?

“Solar radiation” standards describe tests that are designed to characterize the performance of electronics subject to outdoor use or other harsh environments. The two most important solar radiation standards are MIL-STD-810G and IEC 60068 2 5. The Q-SUN product line can be used to meet these environmental “solar radiation” tests. A critical fact to note is that these standards are not specific test methods, despite having sections called “test methods.” Both standards spend considerable space discussing the concept of “tailoring”, which gives laboratories and engineers flexibility in designing tests that apply the environmental stresses discussed in each section. Thus not only are these standards performance-based in allowing multiple hardware designs, it is also possible to alter the actual test conditions if the result meets the general intent of the section. MIL-STD-810H clearly states:

“It is important to note that this document does not impose design or test specifications. Rather, it describes the environmental tailoring process that results in realistic materiel designs and test methods based on materiel system performance requirements.”

MIL-STD-810H and IEC 60068-2-5 both include numerous statements reinforcing this flexibility in selecting test conditions. In fact, the “solar radiation” tests in these standards are impossible to run in any chamber because they a table showing a target spectral power distribution from 280 to 3000 nm, that no artificial light source can meet. Because no single light source actually meets the specification, the user of the standard must apply sound and reasonable engineering principles to tailor the test by defining how the relevant environmental stresses will be applied. Q-Lab has prepared two special letters confirming that Q-SUN xenon-arc weathering testers can meet the performance requirements of Method 505.7 of standard MIL-STD-810H, provided that the document’s guidance on “test tailoring” is followed. Laboratories who need to add these standards to their scope of accreditation need to write a procedure that addresses specifically what test tailoring they have selected for their Q-SUN.

To read more about Q-Lab and to find more articles like this, please visit Q-Lab Blog.
 

FAQ's

Product FAQ's


Contact Us

  1.  

Q-Lab Phone Numbers

USA: +1-440-835-8700
Florida & Arizona Test Services: +1-305-245-5600
UK/Europe: +44-1204-861616
Germany: +49-681-857470
China: +86-21-5879-7970

 
View Full Site